Your browser doesn't support javascript.
loading
Development of validated liquid chromatographic method for estimation of levocetirizine from pharmaceutical dosage forms.
Article Dans Anglais | IMSEAR | ID: sea-151034
ABSTRACT
A simple and rapid high-performance liquid chromatographic (HPLC) method for the determination levocetirizine has been developed. The chromatographic system consisted of a Water 2695 binary gradient pump, Water 2487 dual wavelength absorbance detector, and Empower 2 software. Separation was achieved on the XTerra symmetry C18 column at room temperature. The sample was introduced through an injector valve with a 20 μL sample loop. The results obtained showed a good agreement with the declared content. Recovery values of levocetirizine in tablets were from 99.57-100.48 %. The proposed method is rapid, accurate and selective; it may be used for the quantitative analysis of levocetirizine from raw materials, in bulk drugs and other dosage formulations.

Texte intégral: Disponible Indice: IMSEAR (Asie du Sud-Est) langue: Anglais Année: 2011 Type: Article