Establishment of Thromboelastography reference interval for healthy adults in Hainan / 中国输血杂志
Chinese Journal of Blood Transfusion
; (12): 441-443, 2022.
Article
de Zh
| WPRIM
| ID: wpr-1004286
Bibliothèque responsable:
WPRO
ABSTRACT
【Objective】 To establish the reference interval of Thromboelastography(TEG)of healthy adult in Hainan, compare with the interval provided by the manufacturers, and analyze the influencing factors. 【Methods】 A total of 308 healthy adult volunteers were included according to the inclusion and exclusion criteria. The TEG was analyzed based on adequate quality control, and the reference interval of each parameter was calculated with 95% CI. 【Results】 The reference interval of these volunteers were R 4.27~9.20 min, K 1.60~3.83 min, Angle 44.22°~67.78°, MA 47.82~64.17 mm, CI -5.63~1.12 and G 4.58~9.03. 14.94% (46/308) of these volunteers had at least one index exceeded the reference interval provided by manufacturers. A total of 74 healthy volunteers were diagnosed with coagulation disorder, with the specificity at 79.22%. Significant differences of R, K, Angle, MA, CI, G were observed between males and females (P<0.05). Hypercoagulability was not associated with the age. 【Conclusion】 This study established a reference interval of TEG of healthy adults in Hainan, which provided reference for related clinical and basic science study.
Texte intégral:
1
Indice:
WPRIM
langue:
Zh
Texte intégral:
Chinese Journal of Blood Transfusion
Année:
2022
Type:
Article