Your browser doesn't support javascript.
loading
Comparison of Corneal Thickness and Anterior Chamber Depth Measured With Orbscan, Pentacam, and Ultrasound Pachymetry
Journal of the Korean Ophthalmological Society ; : 664-669, 2009.
Article Dans Coréen | WPRIM | ID: wpr-111149
ABSTRACT

PURPOSE:

We compared the measurements of corneal thickness and anterior chamber depth (ACD) using three different methods Orbscan, Pentacam and ultrasound pachymetry.

METHODS:

In healthy volunteers, central corneal thickness was measured with Orbscan, Pentacam and ultrasound pachymetry. Estimation of peripheral corneal thickness and ACD were done by Orbscan and Pentacam. All results were compared statistically.

RESULTS:

The mean central corneal thickness of 64 eyes measured by Orbscan, Pentacam and ultrasound pachymetry were 509.1+/-61.0 microm, 539.2+/-51.7 microm and 527.4+/-48.3 microm, respectively. There were statistically significant differences in the results among the three methods (repeated-measures analysis of variance (ANOVA), p0.90, p<0.05). The temporal and nasal peripheral corneal thicknesses were thinner in the Pentacam than in the Orbscan (paired t-test, p<0.05). The superior and inferior corneal thickness and ACD were not significantly different.

CONCLUSIONS:

The results of the three instruments have significant correlations with each other, so all methods are clinically useful.
Sujets)

Texte intégral: Disponible Indice: WPRIM (Pacifique occidental) Sujet Principal: Science des ultrasons / Oeil / Chambre antérieure du bulbe oculaire langue: Coréen Texte intégral: Journal of the Korean Ophthalmological Society Année: 2009 Type: Article

Documents relatifs à ce sujet

MEDLINE

...
LILACS

LIS

Texte intégral: Disponible Indice: WPRIM (Pacifique occidental) Sujet Principal: Science des ultrasons / Oeil / Chambre antérieure du bulbe oculaire langue: Coréen Texte intégral: Journal of the Korean Ophthalmological Society Année: 2009 Type: Article