Your browser doesn't support javascript.
loading
Quantitation of meloxicam in the plasma of koalas (Phascolarctos cinereus) by improved high performance liquid chromatography
J. vet. sci ; J. vet. sci;: 7-14, 2013.
Article de En | WPRIM | ID: wpr-142095
Bibliothèque responsable: WPRO
ABSTRACT
An improved method to determine meloxicam (MEL) concentrations in koala plasma using reversed phase high performance liquid chromatography equipped with a photo diode array detector was developed and validated. A plasma sample clean-up step was carried out with hydrophilic-lipophilic copolymer solid phase extraction cartridges. MEL was separated from an endogenous interference using an isocratic mobile phase [acetonitrile and 50 mM potassium phosphate buffer (pH 2.15), 45:55 (v:v)] on a Nova-Pak C18 4-microm (300 x 3.9 mm) column. Retention times for MEL and piroxicam were 8.03 and 5.56 min, respectively. Peak area ratios of MEL to the internal standard (IS) were used for regression analysis of the calibration curve, which was linear from 10 to 1,000 ng/mL (r2 > 0.9998). Average absolute recovery rates were 91% and 96% for MEL and the IS, respectively. This method had sufficient sensitivity (lower quantitation limit of 10 ng/mL), precision, accuracy, and selectivity for routine analysis of MEL in koala plasma using 250-microL sample volumes. Our technique clearly resolved the MEL peak from the complex koala plasma matrix and accurately measured MEL concentrations in small plasma volumes.
Sujet(s)
Mots clés
Texte intégral: 1 Indice: WPRIM Sujet Principal: Contrôle de qualité / Thiazines / Thiazoles / Structure moléculaire / Piroxicam / Anti-inflammatoires non stéroïdiens / Reproductibilité des résultats / Chromatographie en phase liquide à haute performance / Sensibilité et spécificité / Phascolarctidae Type d'étude: Diagnostic_studies Limites du sujet: Animals langue: En Texte intégral: J. vet. sci Année: 2013 Type: Article
Texte intégral: 1 Indice: WPRIM Sujet Principal: Contrôle de qualité / Thiazines / Thiazoles / Structure moléculaire / Piroxicam / Anti-inflammatoires non stéroïdiens / Reproductibilité des résultats / Chromatographie en phase liquide à haute performance / Sensibilité et spécificité / Phascolarctidae Type d'étude: Diagnostic_studies Limites du sujet: Animals langue: En Texte intégral: J. vet. sci Année: 2013 Type: Article