Your browser doesn't support javascript.
loading
Comparison of IOLMaster(R) and A-Scan Ultrasound: Change in Axial Length After Vitrectomy in Macular Disease
Journal of the Korean Ophthalmological Society ; : 1226-1231, 2009.
Article Dans Coréen | WPRIM | ID: wpr-144217
ABSTRACT

PURPOSE:

To evaluate the differences between IOLMaster(R) and A-scans in changes in axial length after vitrectomy in patients with macular disease.

METHODS:

Using IOLMaster(R) and A-scans, we measured preoperative and postoperative axial length in 12 eyes with epiretinal membranes (ERM) and in 8 eyes with macular holes (MH). The relationship between the absolute error in axial length after vitrectomy and both methods was assessed using Mann-Whitney U test. The correlation to central macular thickness was evaluated by Spearman's correlation coefficient.

RESULTS:

In eyes with ERM and MH, preoperative and postoperative axial lengths obtained with both methods had no significant difference (p>0.05). The absolute error in axial length after vitrectomy was not significant using IOLMaster(R) (ERM 0.07+/-0.05 mm, MH 0.04+/-0.02 mm, p>0.05) but was significant using A-scan (ERM 0.20+/-0.11 mm, MH 0.30+/-0.07 mm, p0.05, MH; correlation coefficient = -0.054, p>0.05, A-scan ERM; correlation coefficient = -0.210, p>0.05, MH; correlation coefficient = -0.156, p>0.05).

CONCLUSIONS:

The IOLMaster(R) is more useful than the A-scan when measuring axial length without refractive errors after vitrectomy in eyes with macular disease.
Sujets)

Texte intégral: Disponible Indice: WPRIM (Pacifique occidental) Sujet Principal: Troubles de la réfraction oculaire / Perforations de la rétine / Vitrectomie / Membrane épirétinienne / Oeil Limites du sujet: Humains langue: Coréen Texte intégral: Journal of the Korean Ophthalmological Society Année: 2009 Type: Article

Documents relatifs à ce sujet

MEDLINE

...
LILACS

LIS

Texte intégral: Disponible Indice: WPRIM (Pacifique occidental) Sujet Principal: Troubles de la réfraction oculaire / Perforations de la rétine / Vitrectomie / Membrane épirétinienne / Oeil Limites du sujet: Humains langue: Coréen Texte intégral: Journal of the Korean Ophthalmological Society Année: 2009 Type: Article