Volumetric accuracy of cone-beam computed tomography
Imaging Science in Dentistry
;
: 165-174, 2017.
Article
Dans Anglais
| WPRIM
| ID: wpr-157678
ABSTRACT
PURPOSE:
This study was performed to investigate the influence of object shape and distance from the center of the image on the volumetric accuracy of cone-beam computed tomography (CBCT) scans, according to different parameters of tube voltage and current. MATERIALS ANDMETHODS:
Four geometric objects (cylinder, cube, pyramid, and hexagon) with predefined dimensions were fabricated. The objects consisted of Teflon-perfluoroalkoxy embedded in a hydrocolloid matrix (Dupli-Coe-Loid TM; GC America Inc., Alsip, IL, USA), encased in an acrylic resin cylinder assembly. An Alphard Vega Dental CT system (Asahi Roentgen Ind. Co., Ltd, Kyoto, Japan) was used to acquire CBCT images. OnDemand 3D (CyberMed Inc., Seoul, Korea) software was used for object segmentation and image analysis. The accuracy was expressed by the volume error (VE). The VE was calculated under 3 different exposure settings. The measured volumes of the objects were compared to the true volumes for statistical analysis.RESULTS:
The mean VE ranged from −4.47% to 2.35%. There was no significant relationship between an object's shape and the VE. A significant correlation was found between the distance of the object to the center of the image and the VE. Tube voltage affected the volume measurements and the VE, but tube current did not.CONCLUSION:
The evaluated CBCT device provided satisfactory volume measurements. To assess volume measurements, it might be sufficient to use serial scans with a high resolution, but a low dose. This information may provide useful guidance for assessing volume measurements.
Texte intégral:
Disponible
Indice:
WPRIM (Pacifique occidental)
Sujet Principal:
Amériques
/
Colloïdes
/
Imagerie tridimensionnelle
/
Tomodensitométrie à faisceau conique
/
Séoul
Pays comme sujet:
Asie
langue:
Anglais
Texte intégral:
Imaging Science in Dentistry
Année:
2017
Type:
Article
Documents relatifs à ce sujet
MEDLINE
...
LILACS
LIS