ABSTRACT
PURPOSE:
The aim of this study was to evaluate that a surgeon can safely remove all sonographic evidence of masses in the breast grater than 3.0 cm in greatest dimension using the 8 g hand held Mammotome (MT).METHODS:
From Jan. 2003 to Mar. 2005, a total of 1,368 US-guided MT excision were performed in 1,112 patients at Kangnam Cha hospital. Of these 1,368 lesions 28 lesions with BI-RADS category 3 features by ultrasonography were included in this study. Ultrasonographic follow-up were performed on 3~6 months later to assess residual tissue and scarring.RESULTS:
Mean patient age was 32.0 years (range 20~55 years). The average size of lesion was 3.5 cm (SD+-0.43 cm). All of the lesions were palpable and all of the specimens were benign. Most common pathologic features were fibroadenoma (75.0%) and breast abscess (14.3%). Mean time required to perform mammotome procedures was 12.2+/-8.2 minutes and mean number of cores removed were 35.1+/-30.2 pieces. No bleeding or infections occurred postoperatively and most complications were mild and anticipated.CONCLUSION:
This study demonstrates that percutaneous removal of big breast benign mass above 3 cm in diameter using The MT system is feasible, effective and safe method for the therapeutic management with minimal morbidity without any additional procedures.
Texte intégral:
Disponible
Indice:
WPRIM (Pacifique occidental)
Sujet Principal:
Région mammaire
/
Études de suivi
/
Échographie
/
Cicatrice
/
Fibroadénome
/
Abcès
/
Main
/
Hémorragie
Type d'étude:
Etude diagnostique
/
Étude observationnelle
/
Étude pronostique
Limites du sujet:
Humains
langue:
Coréen
Texte intégral:
Journal of the Korean Surgical Society
Année:
2006
Type:
Article
Documents relatifs à ce sujet
MEDLINE
...
LILACS
LIS