Your browser doesn't support javascript.
loading
Assessment of Attenuation Correction Techniques with a 137Cs Point Source / 대한핵의학회잡지
Korean Journal of Nuclear Medicine ; : 57-68, 2005.
Article Dans Coréen | WPRIM | ID: wpr-182284
ABSTRACT

PURPOSE:

The objective of this study was to assess attenuation correction algorithms with the 137Cs point source for the brain positron emission tomography (PET) imaging process. MATERIALS AND

METHODS:

Four different types of phantoms were used in this study for testing various types of the attenuation correction techniques. Transmission data of a 137Cs point source were acquired after infusing the emission source into phantoms and then the emission data were subsequently acquired in 3D acquisition mode. Scatter corrections were performed with a background tail-fitting algorithm. Emission data were then reconstructed using iterative reconstruction method with a measured (MAC), elliptical (ELAC), segmented (SAC) and remapping (RAC) attenuation correction, respectively. Reconstructed images were then both qualitatively and quantitatively assessed. In addition, reconstructed images of a normal subject were assessed by nuclear medicine physicians. Subtracted images were also compared.

RESULTS:

ELAC, SAC, and RAC provided a uniform phantom image with less noise for a cylindrical phantom. In contrast, a decrease in intensity at the central portion of the attenuation map was noticed at the result of the MAC. Reconstructed images of Jaszack and Hoffan phantoms presented better quality with RAC and SAC. The attenuation of a skull on images of the normal subject was clearly noticed and the attenuation correction without considering the attenuation of the skull resulted in artificial defects on images of the brain.

CONCLUSION:

the complicated and improved attenuation correction methods were needed to obtain the better accuracy of the quantitative brain PET images.
Sujets)

Texte intégral: Disponible Indice: WPRIM (Pacifique occidental) Sujet Principal: Crâne / Encéphale / Tomographie par émission de positons / Bruit / Médecine nucléaire langue: Coréen Texte intégral: Korean Journal of Nuclear Medicine Année: 2005 Type: Article

Documents relatifs à ce sujet

MEDLINE

...
LILACS

LIS

Texte intégral: Disponible Indice: WPRIM (Pacifique occidental) Sujet Principal: Crâne / Encéphale / Tomographie par émission de positons / Bruit / Médecine nucléaire langue: Coréen Texte intégral: Korean Journal of Nuclear Medicine Année: 2005 Type: Article