Effect of Transverse Magnetic Field on Dose Distribution of High Energy Electron Beam / 의학물리
Korean Journal of Medical Physics
;
: 209-213, 2007.
Article
Dans Coréen
| WPRIM
| ID: wpr-213244
ABSTRACT
In this work we have measured the dose distribution and the percent depth dose of 20 MeV electron beam using the X-OMAT films in order to verify the effects of transverse magnetic field on high energy elecrtron beam in a phantom. The result shows about 30% increase of the percent depth dose at 4.5 cm depth under the transverse magnetic field of 1.5 Tesla at 7.5 cm depth. We have verified that these were in an agreement with other theoretical results.