Delayed Type III Endoleak Caused by Fabric Erosion after Endovascular Repair of an Abdominal Aortic Aneurysm
The Korean Journal of Thoracic and Cardiovascular Surgery
;
: 291-294, 2017.
Article
Dans Anglais
| WPRIM
| ID: wpr-217609
ABSTRACT
A 74-year-old patient presented with recurrent aneurysms in the infrarenal abdominal aorta and right common iliac artery 6 years after endovascular aortic repair using endografts in the same location. The patient underwent an aorto-bi-iliac replacement with removal of the stent graft. Two holes measuring 2 mm each were found in the removed graft, and they appeared to have been caused by wear from continuous friction between the endograft and the aortic wall.