Comparison of the Stability Between Three-piece and Single-piece Aspheric Intraocular Lenses
Journal of the Korean Ophthalmological Society
;
: 1584-1589, 2010.
Article
Dans Coréen
| WPRIM
| ID: wpr-218851
ABSTRACT
PURPOSE:
The goal of the present study was to compare the stability of the 3-piece (Tecnis ZA9003) and single-piece (AcrySof IQ) aspheric intraocular lenses (IOL) by testing decentration, tilt, anterior chamber depth, and refraction.METHODS:
The subjects of this study consisted of 101 eyes who had undergone cataract surgeries with a 3-piece aspheric IOL (54 eyes) and with a single-piece aspheric IOL (47 eyes). The decentration, tilt, anterior chamber depth, and refraction were measured on postoperative day 1, 1 month, and 2 months, using an anterior eye segment analysis system (EAS-1000, Nidek, Japan).RESULTS:
There was a statistically significant difference in the decentration on the postoperative day 1 (p = 0.04). However, there was no statistically significant difference on postoperative 1 month (p = 0.15) and 2 months (p = 0.13). There was no statistically significant difference in the tilt on postoperative day 1, 1 month, and 2 months. There was no statistically significant difference in the anterior chamber depth on postoperative day 1, 1 month, and 2 months. There was a statistically significant difference in the refraction on postoperative day 1 (p = 0.03). However, there was no statistically significant difference on postoperative 1 month (p = 0.07) and 2 months (p = 0.07).CONCLUSIONS:
There was no statistically significant difference in the decentration, anterior chamber depth, and refraction between the 3-piece and single-piece aspheric IOL. Therefore, there is no difference between the 3-piece and single-piece aspheric IOL in the capsular bag stability until 2 months postoperatively.
Texte intégral:
Disponible
Indice:
WPRIM (Pacifique occidental)
Sujet Principal:
Cataracte
/
Oeil
/
Lentilles intraoculaires
/
Chambre antérieure du bulbe oculaire
/
Pôle antérieur du bulbe oculaire
langue:
Coréen
Texte intégral:
Journal of the Korean Ophthalmological Society
Année:
2010
Type:
Article
Documents relatifs à ce sujet
MEDLINE
...
LILACS
LIS