Your browser doesn't support javascript.
loading
Comparison of Clinical and Physiological Efficacies of Different Intermittent Sequential Pneumatic Compression Devices in Preventing Deep Vein Thrombosis: A Prospective Randomized Study
Clinics in Orthopedic Surgery ; : 468-475, 2014.
Article Dans Anglais | WPRIM | ID: wpr-223876
ABSTRACT

BACKGROUND:

There are few comparative studies about the optimal method of pneumatic compression to prevent deep vein thrombosis (DVT). The aim of this prospective randomized study was to compare venous hemodynamic changes and their clinical influences between two graded sequential compression groups (an alternate sequential compression device [ASCD] vs. a simultaneous sequential compression device [SSCD]).

METHODS:

In total, 34 patients (68 limbs) undergoing knee and spine operations were prospectively randomized into two device groups (ASCD vs. SSCD groups). Duplex ultrasonography examinations were performed on the 4th and 7th postoperative days for the detection of DVT and the evaluation of venous hemodynamics. Continuous data for the two groups were analyzed using a two-tailed, unpaired t-test. Relative frequencies of unpaired samples were compared using Fisher exact test. Mixed effects models that might be viewed as ANCOVA models were also considered.

RESULTS:

DVT developed in 7 patients (20.6%), all of whom were asymptomatic for isolated calf DVTs. Two of these patients were from the ASCD group (11.8%) and the other five were from the SSCD group (29.4%), but there was no significant difference (p = 0.331). Baseline peak velocity, mean velocity, peak volume flow, and total volume flow were enhanced significantly in both device groups (p < 0.001). However, the degrees of flow and velocity enhancement did not differ significantly between the groups. The accumulated expelled volumes for an hour were in favor of the ASCD group.

CONCLUSIONS:

Both graded sequential compression devices showed similar results both in clinical and physiological efficacies. Further studies are required to investigate the optimal intermittent pneumatic compression method for enhanced hemodynamic efficacy and better thromboprophylaxis.
Sujets)

Texte intégral: Disponible Indice: WPRIM (Pacifique occidental) Sujet Principal: Arthrodèse vertébrale / Rachis / Études prospectives / Facteurs de risque / Résultat thérapeutique / Arthroplastie prothétique de genou / Thrombose veineuse / Dispositifs à compression pneumatique intermittente / Ostéosynthèse / Hémodynamique Type d'étude: Essai clinique contrôlé / Etude d'étiologie / Étude observationnelle / Étude pronostique / Facteurs de risque Limites du sujet: Adulte très âgé / Humains langue: Anglais Texte intégral: Clinics in Orthopedic Surgery Année: 2014 Type: Article

Documents relatifs à ce sujet

MEDLINE

...
LILACS

LIS

Texte intégral: Disponible Indice: WPRIM (Pacifique occidental) Sujet Principal: Arthrodèse vertébrale / Rachis / Études prospectives / Facteurs de risque / Résultat thérapeutique / Arthroplastie prothétique de genou / Thrombose veineuse / Dispositifs à compression pneumatique intermittente / Ostéosynthèse / Hémodynamique Type d'étude: Essai clinique contrôlé / Etude d'étiologie / Étude observationnelle / Étude pronostique / Facteurs de risque Limites du sujet: Adulte très âgé / Humains langue: Anglais Texte intégral: Clinics in Orthopedic Surgery Année: 2014 Type: Article