Radiation Dose Accuracy at the Isocenter: Standard Stereotactic Radiosurgery Technique Developed at Seoul National University Hospital / 대한방사선종양학회지
The Journal of the Korean Society for Therapeutic Radiology and Oncology ; : 391-395, 2002.
Article Dans Coréen | WPRIM | ID: wpr-227233
ABSTRACT

PURPOSE:

To confirm the accuracy of the radiation dose at the isocenter by the standard linear accelerator-based stereotactic radiosurgery technique which was developed at Seoul National University Hospital. MATERIALS AND

METHODS:

Radiation dosimetry was undertaken during standard 5-arc radiosurgery using 6 MV X-ray beam from CL2100C linac. The treatment head was attached with circular tertiary collimators of 10 and 20 mm diameter. We measured the absorbed dose at the isocenter of a multi-purpose phantom using two kinds of detector a 0.125 cc ionization chamber and a silicon diode detector.

RESULTS:

The dose differences at each arc plane between the planned dose and the measured dose at the isocenter raged from -0.73% to -2.69% with the 0.125 cc ion chamber, and from -1.29% to -2.91% with the diode detector during radiosurgery with the tertiary collimator of 20 mm diameter. Those with the 10-mm tertiary collimator ranged from -2.39% to -4.25% with the diode.

CONCLUSION:

The dose accuracy at the isocenter was +/-3%. Therefore, further efforts such as modification in processing of the archived image through DICOM3.0 format are required to lessen the dose difference.