Confirmation of pipetting performances for fully-open automatic biochemistry analyzers / 中国医疗器械杂志
Chinese Journal of Medical Instrumentation ; (6): 277-282, 2005.
Article Dans Chinois | WPRIM | ID: wpr-232956
ABSTRACT
This paper introduces a kind of evaluation method in pipetting performance on new fully automated biochemistry analyzers by experiments. The performance of sample pipetting volume is confirmed by dye dilution method, the performance of reagent pipetting volume and dummy volume is done by weighing method. Meanwhile, researches and comparative researches on dummy volumes in different conditions have been made, providing valuable reference for clinical applications of automatic biochemistry analyzers.
Sujets)
Texte intégral: Disponible Indice: WPRIM (Pacifique occidental) Sujet Principal: Biochimie / Laboratoire automatique / Méthodes langue: Chinois Texte intégral: Chinese Journal of Medical Instrumentation Année: 2005 Type: Article