Your browser doesn't support javascript.
loading
Study on Intelligent Automatic Tracking Radiation Protection Curtain / 中国医疗器械杂志
Chinese Journal of Medical Instrumentation ; (6): 344-346, 2015.
Article Dans Chinois | WPRIM | ID: wpr-265623
ABSTRACT
In order to overcome the shortcomings of traditional X-ray inspection taking passive protection mode, this paper combines the automatic control technology, puts forward a kind of active protection X-ray equipment. The device of automatic detection of patients receiving X-ray irradiation part, intelligent adjustment in patients and shooting device between automatic tracking radiation protection device height. The device has the advantages of automatic adjustment, anti-radiation device, reduce the height of non-irradiated area X-ray radiation and improve the work efficiency. Testing by the professional organization, the device can decrease more than 90% of X-ray dose for patients with non-irradiated area.
Sujets)
Texte intégral: Disponible Indice: WPRIM (Pacifique occidental) Sujet Principal: Dose de rayonnement / Radioprotection / Rayons X / Méthodes Limites du sujet: Humains langue: Chinois Texte intégral: Chinese Journal of Medical Instrumentation Année: 2015 Type: Article