Dosimetric Study of Biliary Stent Loaded with RadioactiveI Seeds / 中华医学杂志(英文版)
Chinese Medical Journal ; (24): 1093-1099, 2017.
Article Dans Anglais | WPRIM | ID: wpr-266856
ABSTRACT
<p><b>BACKGROUND</b>A novel radioactive 125I seed-loaded biliary stent has been used for patients with malignant biliary obstruction. However, the dosimetric characteristics of the stents remain unclear. Therefore, we aimed to describe the dosimetry of the stents of different lengths - with different number as well as activities of 125I seeds.</p><p><b>METHODS</b>The radiation dosimetry of three representative radioactive stent models was evaluated using a treatment planning system (TPS), thermoluminescent dosimeter (TLD) measurements, and Monte Carlo (MC) simulations. In the process of TPS calculation and TLD measurement, two different water-equivalent phantoms were designed to obtain cumulative radial dose distribution. Calibration procedures using TLD in the designed phantom were also conducted. MC simulations were performed using the Monte Carlo N-Particle eXtended version 2.5 general purpose code to calculate the radioactive stent's three-dimensional dose rate distribution in liquid water. Analysis of covariance was used to examine the factors influencing radial dose distribution of the radioactive stent.</p><p><b>RESULTS</b>The maximum reduction in cumulative radial dose was 26% when the seed activity changed from 0.5 mCi to 0.4 mCi for the same length of radioactive stents. The TLD's dose response in the range of 0-10 mGy irradiation by 137Cs γ-ray was linear y = 182225x - 6651.9 (R2=0.99152; y is the irradiation dose in mGy, x is the TLDs' reading in nC). When TLDs were irradiated by different energy radiation sources to a dose of 1 mGy, reading of TLDs was different. Doses at a distance of 0.1 cm from the three stents' surface simulated by MC were 79, 93, and 97 Gy.</p><p><b>CONCLUSIONS</b>TPS calculation, TLD measurement, and MC simulation were performed and were found to be in good agreement. Although the whole experiment was conducted in water-equivalent phantom, data in our evaluation may provide a theoretical basis for dosimetry for the clinical application.</p>
Sujets)
Texte intégral: Disponible Indice: WPRIM (Pacifique occidental) Sujet Principal: Radiométrie / Dosimétrie par thermoluminescence / Simulation numérique / Curiethérapie / Méthode de Monte Carlo / Méthodes Type d'étude: Évaluation en économique de la santé / Étude pronostique Limites du sujet: Humains langue: Anglais Texte intégral: Chinese Medical Journal Année: 2017 Type: Article

Documents relatifs à ce sujet

MEDLINE

...
LILACS

LIS

Texte intégral: Disponible Indice: WPRIM (Pacifique occidental) Sujet Principal: Radiométrie / Dosimétrie par thermoluminescence / Simulation numérique / Curiethérapie / Méthode de Monte Carlo / Méthodes Type d'étude: Évaluation en économique de la santé / Étude pronostique Limites du sujet: Humains langue: Anglais Texte intégral: Chinese Medical Journal Année: 2017 Type: Article