The measurement of retentive force of laser-melted and cast post-keepers of Magfit magnetic attachments / 华西口腔医学杂志
West China Journal of Stomatology
;
(6): 386-388, 2003.
Article
Dans Chinois
| WPRIM
| ID: wpr-283570
ABSTRACT
<p><b>OBJECTIVE</b>In order to compare the influence of casting and laser welding on magnetic retentive force, the magnetic retentive forces of laser-melted and cast post-keepers of the magnetic attachments were measured in vitro.</p><p><b>METHODS</b>5 cast and 6 laser-melted post-keeper specimens (Magfit MD800) were formed, respectively. Using untreated keeper as control group, the vertical magnetic retentive forces were measured by a universal testing machine.</p><p><b>RESULTS</b>The average vertical magnetic retentive force of the laser-melted post-keeper group and the cast post-keeper group in vitro were (4.223 +/- 0.217,0)N, (3.792,9 +/- 0.296,4)N, respectively. Compared with the cast post-keeper group, the average vertical magnetic retentive force of the laser-melted post-keeper groups was much higher in vitro (P < 0.05).</p><p><b>CONCLUSION</b>Laser welding had less influence on the magnetic retentive force of the magnetic attachments than casting.</p>
Texte intégral:
Disponible
Indice:
WPRIM (Pacifique occidental)
Sujet Principal:
Soudage dentaire
/
Acier inoxydable
/
Soudage
/
Technique de coulée dentaire
/
Restauration coronoradiculaire
/
Rétention de prothèse dentaire
/
Alliage dentaire
/
Revêtement coulée dentaire
/
Attachement de précision
/