Your browser doesn't support javascript.
loading
A new parameter measurement system for electrosurgery output / 生物医学工程学杂志
Journal of Biomedical Engineering ; (6): 421-425, 2014.
Article Dans Chinois | WPRIM | ID: wpr-290742
ABSTRACT
Accurate measurements of voltage and current from electrosurgery are the basis of development of electrosurgery with feedback function. We, therefore, developed a parameter measurement system based on PC, with high voltage and current from electrosurgery being sensed with transformers, amplified, filtered, transformed into single-ended signals, and then into RMS signals. The root mean square (RMS) signals were transformed into digital signals through DAQ card and the data was processed in PC with Labview. The process included sampling, displaying and storage. The experiment results indicated that the measurement system could measure the output parameters from electrosurgery steadily and correctly so that the development of the system has been successful. It can be the basis of development of embedded parameters measurement system and can provide accurate feedback information for intellectual electrosurgery.
Sujets)
Texte intégral: Disponible Indice: WPRIM (Pacifique occidental) Sujet Principal: Électrochirurgie / Conception d'appareillage langue: Chinois Texte intégral: Journal of Biomedical Engineering Année: 2014 Type: Article

Documents relatifs à ce sujet

MEDLINE

...
LILACS

LIS

Texte intégral: Disponible Indice: WPRIM (Pacifique occidental) Sujet Principal: Électrochirurgie / Conception d'appareillage langue: Chinois Texte intégral: Journal of Biomedical Engineering Année: 2014 Type: Article