Your browser doesn't support javascript.
loading
Liquid gastric-emptying measurement using an electrical bio-impedance method / 中国医疗器械杂志
Chinese Journal of Medical Instrumentation ; (6): 253-256, 2008.
Article Dans Chinois | WPRIM | ID: wpr-309602
ABSTRACT
This paper introduces a 4-electrode gastric-emptying measurement system using an electrical bio-impedance method by which gastric-emptying experiments of 24 healthy volunteers have been carried out. The test results show that the base impedance is stable, the average GET/2 is 8.78 +/- 1.76 min. The new system provides a new non-invasive measurement method for the clinical study on gastric-emptying functions.
Sujets)
Texte intégral: Disponible Indice: WPRIM (Pacifique occidental) Sujet Principal: Physiologie / Impédance électrique / Consommation alimentaire / Vidange gastrique Limites du sujet: Humains langue: Chinois Texte intégral: Chinese Journal of Medical Instrumentation Année: 2008 Type: Article