Development and research of temporary demand pacemaker with electrocardiosignal display / 生物医学工程学杂志
Journal of Biomedical Engineering
;
(6): 650-653, 2004.
Article
Dans Chinois
| WPRIM
| ID: wpr-342643
ABSTRACT
A temporary demand pacemaker with electrocardiosignal display is introduced in this paper. Double way low-noise electrocardiosignal preamplifier, amplitude limiter, high and low pass filter, 50 Hz notch filter, TTL level generator and stimulating pulse formation circuit are components of the hardware electrocircuit. The demand pacing and the electrocardiosignal display are separately controlled by the software in which the double microcontrollers communications technique is used. In this study, liquid crystal display is firstly used in body surface electrocardiosignal display or intracardial electrophysiologic signal display when the temporary demand pacemaker is installed and put into use. The machine has proven clinically useful and can be of wide appliation.