Design of a microfluidic immunoassay system based on superparamagnetic microbeads / 中国医疗器械杂志
Chinese Journal of Medical Instrumentation ; (6): 369-371, 2006.
Article Dans Chinois | WPRIM | ID: wpr-355376
ABSTRACT
We use direct-write laser micromachining technology to fabricate the microfluidic chip, and to establish a microfluidic chemiluminescence immunoassay system based on superparamagnetic microbeads, for detecting alpha- fetoprotein (AFP). The AFP analysis can be completed in 20 minutes with 5 microl sample and 5 microl reagent, and there is a good linear correlation in the range of 1-800 ng/ml.
Sujets)
Texte intégral: Disponible Indice: WPRIM (Pacifique occidental) Sujet Principal: Dosage immunologique / Alphafoetoprotéines / Techniques d'analyse microfluidique / Conception d'appareillage / Mesures de luminescence / Méthodes Limites du sujet: Humains langue: Chinois Texte intégral: Chinese Journal of Medical Instrumentation Année: 2006 Type: Article

Documents relatifs à ce sujet

MEDLINE

...
LILACS

LIS

Texte intégral: Disponible Indice: WPRIM (Pacifique occidental) Sujet Principal: Dosage immunologique / Alphafoetoprotéines / Techniques d'analyse microfluidique / Conception d'appareillage / Mesures de luminescence / Méthodes Limites du sujet: Humains langue: Chinois Texte intégral: Chinese Journal of Medical Instrumentation Année: 2006 Type: Article