Rectovaginal Fistula Stage-one Repair Device Based on Magnetic Compression Technique / 生物医学工程学杂志
J. biomed. eng
; Sheng wu yi xue gong cheng xue za zhi;(6): 1096-1099, 2015.
Article
de Zh
| WPRIM
| ID: wpr-357913
Bibliothèque responsable:
WPRO
ABSTRACT
The magnamosis device for stage-one repair of the rectovaginal tistula consists of two arc magnets. Drawing the interrupting thread along the fistula margin via the vaginal side, and pulling the string to arrange the magnets at the fistula base along the long axis of the vagina, we made the magnamosis device automatically clipped to seal the fistula. After removing the threads we kept the device for 2-4 weeks till the natural detachment of it when the compressed tissue in between healed after vascular necrosis. This device utilizing the unique ability of magnamosis to fulfill anastomosis under inflammatory infected state reduces the current high relapse rate and colostomy drawbacks of the conventional rectovaginal neoplasty.
Texte intégral:
1
Indice:
WPRIM
Sujet Principal:
Pression
/
Chirurgie générale
/
Procédures de chirurgie gynécologique
/
Cicatrisation de plaie
/
Fistule rectovaginale
/
Aimants
/
Magnétisme
/
Méthodes
Limites du sujet:
Female
/
Humans
langue:
Zh
Texte intégral:
J. biomed. eng
/
Sheng wu yi xue gong cheng xue za zhi
Année:
2015
Type:
Article