Evaluation of the levels of auto-PEEP in the patients with acute respiratory failure required mechanical ventilation
Journal of Medical Research
;
: 46-51, 2004.
Article
Dans Vietnamien
| WPRIM
| ID: wpr-3607
ABSTRACT
We have studied 27 patients on medical ventilation in whom we measured auto-PEEP by means of end expiratory airway occlusion. Four patients with ARDS had 10.25 2.05 cm H2O of auto PEEP. Six patients with COPD had 12.83 4.98 cm H2O of auto PEEP. Seven patients with acute severe asthma had 20.29 7.21 cm H2O of auto PEEP. Ten patients with neuromuscular disease had 5.2 4.07 cm H2O of auto PEEP. PEEP was set equal to half the auto-PEEP to treat patients with ARDS, COPD and acute severe asthma. We have show that the value of auto PEEP fell to 4.67 3.7 cm H2O in patient with ARDS, to 7 2.31 in patient of COPD and to 9.71 1.67 in patients of asthma. PEEP 50% of auto-PEEP is safe and does not worsen pulmonary hyperinflation
Texte intégral:
Disponible
Indice:
WPRIM (Pacifique occidental)
Sujet Principal:
Ventilation artificielle
/
Insuffisance respiratoire
/
Respiration avec pression positive intrinsèque
langue:
Vietnamien
Texte intégral:
Journal of Medical Research
Année:
2004
Type:
Article
Documents relatifs à ce sujet
MEDLINE
...
LILACS
LIS