Your browser doesn't support javascript.
loading
Quality control and quality assurance of single photon emission computed tomography/computed tomography ( SPECT/CT) system / 中国基层医药
Article de Zh | WPRIM | ID: wpr-386359
Bibliothèque responsable: WPRO
ABSTRACT
Objective To investigate the quality control and quality assurance of the SPECT/CT system.Methods The energy peak,energy resolution capacity and the intrinsic uniformity of the device's detector were regularly examined, and the quality control was performed. Results The daily hardware maintain could reduce the rate of system's trouble. The device's energy peak,energy resolution capacity were consistent during half year's observation period. The two detector's intrinsic uniformity were better after calibration than before. [ detector Ⅰ: ( 2.71 ± 0.28 ) vs (2.37±0.11)(t=2.489,P<0. 05);detector Ⅱ:(2.68 ±0.12)vs(2.38 ±0. 19)(t =6.421,P <0.01) ] .Conclusion Regular quality control and maintain could keep the function stabilization,enhance the availability rate of the SPECT/CT system.
Mots clés
Texte intégral: 1 Indice: WPRIM langue: Zh Texte intégral: Chinese Journal of Primary Medicine and Pharmacy Année: 2010 Type: Article
Texte intégral: 1 Indice: WPRIM langue: Zh Texte intégral: Chinese Journal of Primary Medicine and Pharmacy Année: 2010 Type: Article