Your browser doesn't support javascript.
loading
Three-dimensional dose rate table for 103pd radioactive stent / 中华放射肿瘤学杂志
Article de Zh | WPRIM | ID: wpr-401421
Bibliothèque responsable: WPRO
ABSTRACT
Objective To investigate three-dimensional dose distribution for 103Pd radioactive stent.Methods The surface dose,the axial dose and radial dose in surface for 103pd stent (3 mm × 13 mm) were estimated by experimental simulating method, analytic function and MCNP4b code. Three-dimensional dose distribution was calculated by MCNP4b code. Results The surface dose of 103pd stent was 0. 109 and 0. 106 Gy estimated by experimental simulating method and MCNP4b code,between which the difference was less than 3%. The axial dose calculated by analytic function and MCNP4b code was well consistent,and so was the radial dose estimated by the three methods. Dose rate table were estimated by MCNP4b code. Conclnsions Dose distribution for 103 Pd stent estimated by the three methods is relatively accurate. Three-dimensional dose table estimated by MCNP4b may be used to calculate dose for 103Pd stent in animal experiment and clinical application.
Mots clés
Texte intégral: 1 Indice: WPRIM langue: Zh Texte intégral: Chinese Journal of Radiation Oncology Année: 2008 Type: Article
Texte intégral: 1 Indice: WPRIM langue: Zh Texte intégral: Chinese Journal of Radiation Oncology Année: 2008 Type: Article