Comparison of Magnetic Navigation System and Conventional Method in Catheter Ablation of Atrial Fibrillation: Is Magnetic Navigation System Is More Effective and Safer Than Conventional Method?
Korean Circulation Journal ; : 248-252, 2011.
Article Dans Anglais | WPRIM | ID: wpr-43509
ABSTRACT
BACKGROUND AND

OBJECTIVES:

Although there have been so many reports of catheter ablation of atrial fibrillation (AF) with magnetic navigation system (MNS), it is not necessarily obvious that MNS is more effective than conventional ablation. We performed AF ablation with MNS and compared the clinical outcomes and radiofrequency ablation parameters with those of conventional ablation. SUBJECTS AND

METHODS:

One hundred eleven consecutive patients (conventional group, n=70 vs. MNS group, n=41) undergoing catheter ablation of AF were enrolled. We compared and analyzed the procedural parameters, namely fluoroscopic time, procedural time, acute procedural success and 3 months success rate of both groups.

RESULTS:

The MNS group was associated with slightly larger left atrial size (43.7+/-6.3 mm vs. 41.2+/-6.3 mm, p=0.04), significantly longer total procedure time (352+/-50 minutes vs. 283+/-75 minutes, p<0.0001), and shorter total fluoroscopic time (99+/-28 minutes vs. 238+/-45 minutes, p<0.0001) than the conventional group. The MNS and conventional group did not differ with respect to acute procedural success, AF recurrence, atrial flutter/atrial tachycardia recurrence, or total arrhythmia recurrence. While no complications were observed in the MNS group, eight cases of significant pericardial effusion occurred in the conventional group.

CONCLUSION:

The MNS system seems to be effective and safe in the catheter ablation of AF, particularly in the population of patients with persistent AF and slightly dilated left atria.
Sujets)

Texte intégral: Disponible Indice: WPRIM (Pacifique occidental) Sujet Principal: Épanchement péricardique / Troubles du rythme cardiaque / Récidive / Fibrillation auriculaire / Tachycardie / Ablation par cathéter / Cathéters / Aimants / Magnétisme Limites du sujet: Humains langue: Anglais Texte intégral: Korean Circulation Journal Année: 2011 Type: Article

Documents relatifs à ce sujet

MEDLINE

...
LILACS

LIS

Texte intégral: Disponible Indice: WPRIM (Pacifique occidental) Sujet Principal: Épanchement péricardique / Troubles du rythme cardiaque / Récidive / Fibrillation auriculaire / Tachycardie / Ablation par cathéter / Cathéters / Aimants / Magnétisme Limites du sujet: Humains langue: Anglais Texte intégral: Korean Circulation Journal Année: 2011 Type: Article