Application of Diana automated blood grouping analyzer in blood grouping and cross matching / 国际检验医学杂志
International Journal of Laboratory Medicine
; (12): 1480-1481, 2014.
Article
de Zh
| WPRIM
| ID: wpr-451120
Bibliothèque responsable:
WPRO
ABSTRACT
Objective To evaluate the accuracy and reliability of Diana automated blood grouping analyzer in blood grouping and cross matching.Methods 2 300 patients′ABO and RhD blood groups were examined by conventional tube test and the fully auto-mated blood grouping analyzer and 900 patients′samples were tested using Diana automated blood grouping for blood cross matc-hing,and it was compared with polymatching method.Results The analyzer′s accuracy rate of blood grouping by two methods were 99.87% and 100.00%.The incompatibility occurred in 30 specimens in automatic blood type instrument,in 3 specimens in manual polymatching method.Conclusion The results of Diana automated blood grouping analyzer used for blood grouping and cross matc-hing blood testing are reliable.Its experimental operation is normalized and standardized with an advantage of low incidence of human er-ror.Moreover,the experimental results can be permanently preserved,which provides a convenience to search for medical proof.