Your browser doesn't support javascript.
loading
Content Determination of Valacyclovir Hydrochloride Capsules by HPLC / 中国药师
China Pharmacist ; (12): 238-240, 2014.
Article Dans Chinois | WPRIM | ID: wpr-452780
ABSTRACT

Objective:

To develop an HPLC method for the determination of valacyclovir hydrochloride capsules.

Methods:

An Inertsil ODS (250 mm × 4. 6 mm,5μm) column was employed with methanol-0. 02 mol·L-1 potassium dihydrogen phosphate solution (20∶80) as the mobile phase, the flow rate was 1. 0 ml·min-1, the detection wavelength was at 251 nm, the column temperature was 30℃,and the sample size was 20μl.

Results:

The linear range of valacyclovir hydrochloride was 2-40μg·ml-1 , the correlation coeffi-cient was 0. 999 9 and the average recovery was 99. 8%(RSD=0. 16%,n=9).

Conclusion:

The analytical method is simple, accu-rate and special, which can be used in the content determination and quality control of valacyclovir hydrochloride capsules.

Texte intégral: Disponible Indice: WPRIM (Pacifique occidental) langue: Chinois Texte intégral: China Pharmacist Année: 2014 Type: Article

Documents relatifs à ce sujet