ABSTRACT
Objective To establish a method to quantify the residual cross-linker vinyl sulfone and its hydration product 1 ,4-Thioxane-1 ,1-Dioxide.Methods The polysaccharide was precipitated by ethanol,however vinyl sulfone and its hydration product were soluble,and can be analyzed by direct injection.The analysis was carried on Agilent DB-wax capillary column (30m ×0.53 mm,1.0 μm).The flame ionization detector (FID)and flame photometric detector (FPD)were used to detect samples and the efficiency were compared. Results The linear range of vinyl sulfone and its hydration product were separately 0.5 ~20μg/mL and 2~100μg/mL detected by FID.The Limit of Quantity (LOQ)were 0.8μg/mL and 2.3μg/mL, respctively.The Limit of Detection (LOD)were 2.6μg/mLand 7.6μg/mL,respectively,and the average recoveries (n=9)of them were 104.3%and 92.4%,respectively.Conclusion FID could meet the needs of the test,and this method is simple and accurate with high sensitivity and good repeatability,which can be used for quality control of trace vinyl sulfone and its hydration products 1 ,4-Thioxane-1 ,1-Dioxide in cross-linked sodium hyaluronate injection.
Mots clés
Texte intégral: 1 Indice: WPRIM langue: Zh Texte intégral: Chinese Journal of Biochemical Pharmaceutics Année: 2014 Type: Article