Your browser doesn't support javascript.
loading
Laser Ablation Inductively Coupled Plasma Mass Spectrometric Method for Determination of Cr Content and Distribution in Cr_doped ZnSe Crystals / 分析化学
Article de Zh | WPRIM | ID: wpr-457778
Bibliothèque responsable: WPRO
ABSTRACT
Laser ablation inductively coupled plasma mass spectrometry ( LA_ICP_MS ) was applied for the determination of doping element chromium( Cr) content and distribution in Cr∶ZnSe crystals. Several different Cr∶ZnSe crystals were prepared by diffusion method as reference material to solve the problem of accurate quantization. The homogeneity of Cr in these samples was characterized by LA_ICP_MS and the concentrations achieved by inductively coupled plasma atomic emission spectrometry ( ICP_AES ) . With signal pot and line scan sampling, the present method provided effective position and content distribution information of Cr in ZnSe crystals, achieved the in situ analysis. The correlation coefficient of Cr in calibration curve was 0. 9992 and the detection limit was 0. 08 mg/kg. It could provide effective means for the distribution statistics of doping element in different growth condition crystals.
Mots clés
Texte intégral: 1 Indice: WPRIM langue: Zh Texte intégral: Chinese Journal of Analytical Chemistry Année: 2015 Type: Article