Your browser doesn't support javascript.
loading
Hemostasis effect comparison between artificial pressing and Perclose vascular closure device on femoral artery / 心血管康复医学杂志
Chinese Journal of cardiovascular Rehabilitation Medicine ; (6): 563-565, 2014.
Article Dans Chinois | WPRIM | ID: wpr-473655
ABSTRACT

Objective:

To observe hemostoisis efficacy and safety of Perclose vascular closure device on femoral artery in transfemoral pathway radiofrequency ablation (RFA) of supraventricular tachycardia (SVT) .

Methods:

A total of 199 patients undergoing transfemoral RFA of SVT were divided into Perclose vascular closure group (Perclose group ,n=98) and artificial pressing group (n=101) .Immediate success rate ,hemostasis time ,braking time and in-cidence rates of postoperative complications were observed between two groups .

Results:

There was no significant difference in immediate success rate of hemostasis (98% vs .100% ) between Perclose group and artificial pressing group , P> 0.05. Compared with artificial pressing group , there were significant reductions in hemostasis time [ (18.9 ± 9.1) min vs .(2.7 ± 0.7) min] ,braking time [ (21.6 ± 3.4) min vs .(6.3 ± 2.4) min] and total inci-dence rate of complications (16.8% vs .4.1% ) in Perclose group , P 0.05 all) .

Conclusion:

Perclose vascular closure device could shorten the duration of hemostasis and braking ,and reduce the complications .It could be the preferred choice for femoral artery hemostasis .

Texte intégral: Disponible Indice: WPRIM (Pacifique occidental) langue: Chinois Texte intégral: Chinese Journal of cardiovascular Rehabilitation Medicine Année: 2014 Type: Article

Documents relatifs à ce sujet

MEDLINE

...
LILACS

LIS

Texte intégral: Disponible Indice: WPRIM (Pacifique occidental) langue: Chinois Texte intégral: Chinese Journal of cardiovascular Rehabilitation Medicine Année: 2014 Type: Article