Initial experience on protocol optimization for integrated PET/MR / 中华核医学与分子影像杂志
Chinese Journal of Nuclear Medicine and Molecular Imaging
;
(6): 438-442, 2014.
Article
Dans Chinois
| WPRIM
| ID: wpr-474703
ABSTRACT
Objective To investigate the optimal workflow and protocol for integrated PET/MR by comparison with PET/CT.Methods A total of 228 patients were enrolled in this study for PET/CT and PET/MR evaluation on the same day.Six PET/MR protocols with different MR sequences but the same PET acquisition protocol were investigated and the optimal protocol was identified based on image quality,acquisition time and diagnostic performance.Results PET/MR workflow was similar to PET/CT,however,some special issues needed to be considered for PET/MR.Among the 6 protocols,protocol No.6 outperformed others for body and head regions.Types of artifacts were found more often in PET/MR than in PET/CT.Conclusions By optimizing the protocol,PET/MR could achieve almost the same diagnostic performance as PET/CT.However,the issues of long acquisition time and artifacts on PET/MR need to be further improved.