Your browser doesn't support javascript.
loading
Design of multi-channel dry type biochemistry sensors based on fiber bundles / 军事医学
Military Medical Sciences ; (12): 582-586, 2015.
Article Dans Chinois | WPRIM | ID: wpr-476587
ABSTRACT
Objective To develop a multi-channel dry type biochemistry sensor with a compact structure and high measurement accuracy.Methods The principle of double beam compensation based on reference LED was applied to improve the measurement accuracy.The complex splitting system was replaced by MXN fiber bundle and free-form surface lens to make the instrument more compact and lightweight.Use of the adaptive amplification photoelectric detection improved the measurement accuracy while simplifying the process.Results and Conclusion It has been proved by experiments that this sensor has the advantages of high measurement accuracy, little interference and compact construction. This sensor may well meet the requirements of dry type biochemistry analysis.

Texte intégral: Disponible Indice: WPRIM (Pacifique occidental) langue: Chinois Texte intégral: Military Medical Sciences Année: 2015 Type: Article

Documents relatifs à ce sujet

MEDLINE

...
LILACS

LIS

Texte intégral: Disponible Indice: WPRIM (Pacifique occidental) langue: Chinois Texte intégral: Military Medical Sciences Année: 2015 Type: Article