Performance verification of Hitachi 7600 automatic biochemical analyzer / 国际检验医学杂志
International Journal of Laboratory Medicine
;
(12): 2983-2986, 2015.
Article
Dans Chinois
| WPRIM
| ID: wpr-481913
ABSTRACT
Objective To verify the performance of Hitachi 7600 automatic biochemical analyzer .Methods According to labora‐tory accreditation criteria and the performance verification documents of American CLIA′88 ,the precision ,accuracy ,reference inter‐val ,linear range and clinical reportable range were verified for the tests of 19 conventional biochemical indicators performed on Hita‐chi 7600 biochemical analyzer most often .Results The precision ,accuracy ,reference interval ,linear range and clinical reportable range were all acceptable .Conclusion Hitachi 7600 automatic biochemical analyzer could fully meet the reqirements in clinical ap‐plication .