To master the performances of the cardiac defibrillator-monitor and enhance its reliability and effi-ciency through quality control. ESA612 electrical safety analyzer and Phase3 defibrillator analyzer were used to test the performances of 56 cardiac defibrillator-monitors, involving in basic items detection, electrical safety detection and performance detection. The qualified items included charging times, internal discharge, ECG monitoring and etc, while the unqualified ones included appearance, electrical safety as well as delay time under synchronous mode. The quality control can improve the safety of the cardiac defibrillator-monitor.