Determination of Matrine in Matrine Ethosome Thermosensitive Gels by HPLC / 医药导报
Herald of Medicine
; (12): 521-523, 2016.
Article
de Zh
| WPRIM
| ID: wpr-486531
Bibliothèque responsable:
WPRO
ABSTRACT
Objective To establish an HPLC method for determination of matrine in matrine ethosome thermosensitive gels. Methods Matrine was detected in matrine ethosome thermosensitive gels by using a YMC-Pack NH2 column(4.6 mm× 150 mm,5 μm),with the mobile phase containing acetonitrile-ethanol(9: 1).The flow rate was 0.8 mL·min-1 ,the column temperature( 25 ± 2) ℃ and the UV detection wavelength 215 nm. Results The peak area correlated with matrine concentration within the range of 1.00-100.00 μg·mL-1 linearly,and the average correlation coefficient was 0.999 2.The average recoveries rate of low,medium and high concentration of matrine were 98.16%,98.15%,98.00%,respectively,and the RSD were 0.30%,0. 69%,0. 71%,respectively. Conclusion The method is simple,sensitive and accurate. With the advantages of specificity,high precision,good stability and repeatability,the HPLC method can be used for the determination of matrine in matrine ethosome thermosensitive gels.
Texte intégral:
1
Indice:
WPRIM
langue:
Zh
Texte intégral:
Herald of Medicine
Année:
2016
Type:
Article