Your browser doesn't support javascript.
loading
Design and Implementation of Embedded Telemetry System for Amperometric Detection / 分析化学
Chinese Journal of Analytical Chemistry ; (12): 1465-1470, 2016.
Article Dans Chinois | WPRIM | ID: wpr-503578
ABSTRACT
A new micro embedded telemetry system was developed for the amperometric sensing detection. Its output voltage range was ±0. 5 V and resolution was<1 mV. The current acquisition range was ±1μA and the minimum resolution was 0. 2 nA. This telemetry system was designed based on microprocessor ADuCM360, including a potentiostat, a current detection module and a radio module. And the size was only 24 mm × 13 mm × 11 mm. The computer software written in LabVIEW language was used for data storage and display. In order to verify the accuracy and reliability of this system, an electrical performance test was performed. The current response for ascorbic acid with different concentration was recorded by using the telemetry system. The potential of working electrode was set as 30 mV, and the current response of ascorbic acid electrode had a good linear relationship with its concentration within the concentration range of 50-300 μmol/L. Linear equation was I(nA) = 2. 98CAA(μmol/L)-137. 39, and linear correlation coefficient R2=0. 984. Moreover, the applicability of the instrument in the study for living animals was explored by using the cerebral ischemia as model.

Texte intégral: Disponible Indice: WPRIM (Pacifique occidental) Type d'étude: Etude diagnostique langue: Chinois Texte intégral: Chinese Journal of Analytical Chemistry Année: 2016 Type: Article

Documents relatifs à ce sujet

MEDLINE

...
LILACS

LIS

Texte intégral: Disponible Indice: WPRIM (Pacifique occidental) Type d'étude: Etude diagnostique langue: Chinois Texte intégral: Chinese Journal of Analytical Chemistry Année: 2016 Type: Article