Evaluation of three glucometers for whole blood glucose measurements at the point of care in preterm or low-birth-weight infants / 소아과
Korean Journal of Pediatrics
;
: 301-308, 2015.
Article
Dans Anglais
| WPRIM
| ID: wpr-50471
ABSTRACT
PURPOSE:
We evaluated three blood glucose self-monitoring for measuring whole blood glucose levels in preterm and low-birth-weight infants.METHODS:
Between December 1, 2012 and March 31, 2013, 230 blood samples were collected from 50 newborns, who weighed, 60 mg/dL, the correlation coefficients were 0.822, 0.819, and 0.839, respectively. All correlation coefficients were statistically significant. And the values from the blood glucose self-monitoring were not significantly different from the value of the laboratory equipment , after correcting for each device's average value (P>0.05). When using laboratory equipment (blood glucose < or =60 mg/dL), each device had a sensitivity of 0.458, 0.604, and 0.688 and a specificity of 0.995, 0.989, and 0.989, respectively.CONCLUSION:
Significant difference is not found between three blood glucose self-monitoring and laboratory equipment. But correlation between the measured values from blood glucose self-monitoring and laboratory equipment is lower in preterm or low-birth-weight infants than adults.
Texte intégral:
Disponible
Indice:
WPRIM (Pacifique occidental)
Sujet Principal:
Glycémie
/
Nourrisson à faible poids de naissance
/
Autosurveillance glycémique
/
Soins intensifs néonatals
/
Sensibilité et spécificité
/
Prestations des soins de santé
/
Glucose
/
Hypoglycémie
Type d'étude:
Etude diagnostique
Limites du sujet:
Adulte
/
Humains
/
Nouveau-né
langue:
Anglais
Texte intégral:
Korean Journal of Pediatrics
Année:
2015
Type:
Article
Documents relatifs à ce sujet
MEDLINE
...
LILACS
LIS