Determination of 4-Hydrazinobenzenesulfonamide Hydrochloride in Celecoxib by HPLC / 中国药师
China Pharmacist
;
(12): 360-362, 2017.
Article
Dans Chinois
| WPRIM
| ID: wpr-507554
ABSTRACT
Objective:
To establish a method for the determination of 4-hydrazinobenzenesulfonamide hydrochloride in celecoxib.Methods:
HPLC was used and the chromatographic column was a Hypercarb series C18 column (150 mm × 4. 6 mm, 3 μm) with 100% porous graphite as the filler. The mobile phase consisted of 10 mmol·L-1 ammonium acetate aqueous solution-methanol (5050) and methanol. The flow rate was 0. 8 ml·min-1 and the detection wavelength was 220 nm. The column temperature was 35 ℃and the injection volume was 20 μl.Results:
The linear range was 99. 897-599. 382 ng·ml-1(r=0. 9982) for 4-hydrazinobenzene-sulfonamide hydrochloride. The average recovery was 97. 4%(RSD=3. 0%,n=9).Conclusion:
The method is simple and accurate, and can be used for the determination of 4-hydrazinobenzenesulfonamide hydrochloride in celecoxib.