Your browser doesn't support javascript.
loading
Technological analysis and comparison for dose monitoring and control systems of two types of linear accelerators / 中国医学装备
China Medical Equipment ; (12): 27-33, 2017.
Article de Zh | WPRIM | ID: wpr-512201
Bibliothèque responsable: WPRO
ABSTRACT
Objective:To analyze and compare the dose monitoring and control systems of two types of medical linear accelerator systematically from the view of mechanical and electrical engineering, and establish the method of the quality assurance of the dosimetry system and the solution in the routine maintenance.Methods: To systematically compare and analyze the system of dose monitoring and control for two types of linear accelerators (Precise Elekta and C-series of Varian Medical Systems) from twofold aspects that included machine physics and electric engineering of medical linear accelerators.Results:The monitoring system that took ionization chamber as center was established between connotation of machine physics and concrete realization of electromechanical engineering, and explained the difference and similarity between open type and sealed type of ionization chamber, and the theory that Precise series needn't be serviced by PFN.Conclusion: Based on the analysis of ionization chambers structure of medical linear accelerators on the principle of quality assurance system, proposed dosimetric testing and adjustment method, and provide reference for the establishment of quality assurance system and technical procedures of the maintenance engineering.
Mots clés
Texte intégral: 1 Indice: WPRIM langue: Zh Texte intégral: China Medical Equipment Année: 2017 Type: Article
Texte intégral: 1 Indice: WPRIM langue: Zh Texte intégral: China Medical Equipment Année: 2017 Type: Article