Eatimation of effective doses from a dental cone beam computed tomographic scanner / 中华放射医学与防护杂志
Article de Zh | WPRIM | ID: wpr-515210
Bibliothèque responsable: WPRO
ABSTRACT
Objective To calculate effective dose based on the dose-area Product (DAP) measured in dental cone beam computed tomograhic (CBCT) scanning and using a conversion coefficient.Methods The method of the DAP was used to calculate the effective dose from KODAK 9500 dental CBCT scanner with fields of view (FOV) of 16 cm×15 cm and 12 cmu×8 cm.The DAP-values (mGy · cm2) were measured using a transmission ionization chamber connected to an electrometer (Doseguard 100,RTI Electronics AB,Sweden) and the chamber was placed directly on the exit of the X-ray tube.All measurements were repeated three times and averaged on each FOV data.The conversion coefficient of 0.080 μSv/mGy · cm2 was used to estimate the effective dose on the basis of DAP values,which was quoted from a literature.Results DAP values were found between the two FAOs to vary from 512 to 1 083 mGy · cm2,and the effective doses calculated with 0.080 μSv/mGy · cm2were 41 to 87 μSv.The effective doses given in this paper was lower than those reported in other literatures.Conclusions DAP measurement was found to be well defined and easy-to-use method of determining effective dose for some CBCT Units.However,determination of specific conversion coefficient in the CBCT must be further developed.
Mots clés
Texte intégral: 1 Indice: WPRIM langue: Zh Texte intégral: Chinese Journal of Radiological Medicine and Protection Année: 2017 Type: Article
Texte intégral: 1 Indice: WPRIM langue: Zh Texte intégral: Chinese Journal of Radiological Medicine and Protection Année: 2017 Type: Article