Your browser doesn't support javascript.
loading
Determination of Serum Concentration of Metoprolol by HPLC-Fluorescence Detection Method / 中国药房
China Pharmacy ; (12)2007.
Article Dans Chinois | WPRIM | ID: wpr-531261
ABSTRACT

OBJECTIVE:

To establish a HPLC-fluorescence detection method for the determiniation of the serum concentration of metoprolol.

METHODS:

The determination was performed on Symmetry C18 column,and the mobile phase consisted of acetonitrile-water-triethylamine -phosphoric acid(110∶390∶2.5∶1.6) at a flow rate of 0.8 mL?min-1.The excitation wavelength was 265 nm and the emission wavelength 298 nm,and the sample size was 20 ?L.

RESULTS:

Good linearity was obtained for metoprolol over the range of 2.0~100.0 ?g?L-1 with correlation coefficient r=0.999 6.At low,medium and high concentrations,the average recoveries of metoprolol were 99.73%,98.21% and 99.38%,respectively.The intra-day RSD were 2.89%,2.36% and 1.32%,respectively,and the inter-day RSD were 3.73%,3.03% and 2.25% respectively.The lowest detectable limit was 1.0 ?g?L-1.

CONCLUSION:

This method is precise,accurate,specific,simple yet with high recovery,and it is applicable for clinical monitoring of blood concentration and pharmacokinetic study.

Texte intégral: Disponible Indice: WPRIM (Pacifique occidental) Type d'étude: Etude diagnostique langue: Chinois Texte intégral: China Pharmacy Année: 2007 Type: Article

Documents relatifs à ce sujet

MEDLINE

...
LILACS

LIS

Texte intégral: Disponible Indice: WPRIM (Pacifique occidental) Type d'étude: Etude diagnostique langue: Chinois Texte intégral: China Pharmacy Année: 2007 Type: Article