Bacterial Endotoxin Test of Pulse-Engendering Injection / 中成药
Chinese Traditional Patent Medicine
;
(12)1992.
Article
Dans Chinois
| WPRIM
| ID: wpr-569832
ABSTRACT
Objective:
To perform the bacterial endotoxin test of Pulse-Engendering Injection.Methods:
The interference actions of the different dilutabilities of injection at the greatest effective diluent range were observed with some sensitivity of TAL.Results:
When 0.5EU/ml sensitivity of TAL was used and the dilutability of sample was 1∶8, the injection showed obvious enhancement action, but when 0.25EU/ml senitiity of TAL was used and the dilutability was 1∶16, the enhancement action of injection could be eliminated.Conclusion:
The bacterial endoxin test can be carried out with 0.25EU/ml sensitivity of TAL when the dilutability of sample is 1∶16, the limit dose of bacterial endotoxin is controlled thereby.