Study on the replace time of breathing machine humidification pot liquid in mechanical ventilation / 重庆医科大学学报
Journal of Chongqing Medical University ; (12)2003.
Article Dans Chinois | WPRIM | ID: wpr-581006
ABSTRACT

Objective:

To study on the best replacement time of breathing humidification pot liquid in mechanical ventilation.

Methods:

33 mechanical ventilation patients were randomly divided into observation group and control group,in the observation group the humidification pot liquid were replaced 7 days,while in a control group were replaced every day,all liquid samples were cultured for bacteria.

Results:

119 samples collected in observation group have 3 cases with positive bacterial growth(2.5%),while 112 samples collected in control group have 15 cases with positive bacterial growth(13.4%).

Conclusion:

Replace humidification pot liquid of breathing machine everyday may cause bacterial pollution more easily than they were replaced in 7 days interval.