Your browser doesn't support javascript.
loading
Comparison study between electrodes and ending reaction methods for blood glucose and urea measurement / 中华检验医学杂志
Article de Zh | WPRIM | ID: wpr-582569
Bibliothèque responsable: WPRO
ABSTRACT
Objective To evaluate the reliability and clinical application of measuring blood glucose and urea with electrodes and end reaction methods. Method Using the electrodes and end reaction methods, blood glucose and urea were measured respectively. The results and the relativity between the two methods were compared. Result Comparing the two methods for GLU and UREA, the correlative factors were 0.990 1 and 0.989 1 respectively. They had fairly good comparability. The linear ranges of their regression equations met the clinical needs. With favorable accuracy, the correlative factors were above 0.997 9. Conclusion The electrodes method was rapid and accurate, which was suitable for emergency tests. The ending reaction methods was fairly stable and has less interfering factors, which was fit for the routine use.
Mots clés
Texte intégral: 1 Indice: WPRIM langue: Zh Texte intégral: Chinese Journal of Laboratory Medicine Année: 2001 Type: Article
Texte intégral: 1 Indice: WPRIM langue: Zh Texte intégral: Chinese Journal of Laboratory Medicine Année: 2001 Type: Article