Application of uniformity correction technology of Siemens E.CAM U-fly / 医疗卫生装备
Chinese Medical Equipment Journal
;
(6)2004.
Article
Dans Chinois
| WPRIM
| ID: wpr-587306
ABSTRACT
The efficiency of Siemens E.CAM is evaluated by analyzing and practicing the technology of Siemens U-fly.Basic testing procedure is run.Standard source 99mTc and 57Co is used respectively and the operation is performed according to the instruction.The detecting head's uniformity can be corrected in this way with satisfactory results.It is suggested uniformity should be tested every day before study because it is the most important factor for the quality of clinical images.
Texte intégral:
Disponible
Indice:
WPRIM (Pacifique occidental)
langue:
Chinois
Texte intégral:
Chinese Medical Equipment Journal
Année:
2004
Type:
Article
Documents relatifs à ce sujet
MEDLINE
...
LILACS
LIS