Your browser doesn't support javascript.
loading
Time-domain Photoacoustic Tomography of Mouse Brain Vessel in Vivo / 医疗卫生装备
Article de Zh | WPRIM | ID: wpr-596607
Bibliothèque responsable: WPRO
ABSTRACT
Objective To get photoacoustic (PA) image of vascular structure inside mouse brain in vivo.Methods A 532nm pulsed laser was used as the pumping source to generate PA signal,and a wide-band PVDF unfocused ultrasound transducer was used to collect time-domain PA signal by circular scan.PA signal recorded at each scanning position was then de-noised by wavelets transform using soft-thresholding.The PA image of the mouse brain vessel was reconstructed by time-domain spherical back projection algorithm.Results PA image agreed well with the histological picture of mouse brain vessel.Conclusion The experimental result indicates the potential value of clinical application.
Mots clés
Texte intégral: 1 Indice: WPRIM langue: Zh Texte intégral: Chinese Medical Equipment Journal Année: 2003 Type: Article