Comparison analysis between different analysis systems for the detection of electrolytes / 国际检验医学杂志
International Journal of Laboratory Medicine
; (12): 1108-1109,1112, 2015.
Article
de Zh
| WPRIM
| ID: wpr-601061
Bibliothèque responsable:
WPRO
ABSTRACT
Objective To investigate the comparability of K + ,Na+ and Cl- measurement between Cobas8000 analyzer and AU640 analyzer .Methods According to the EP9‐A2 document of the National Committee for Clinical Laboratory Standards ,the precision of AU640 analyzer for the detection of K + ,Na+ and Cl- was evaluated ,and the correlation between the detected results of the two analyzers were also analyzed .Results AU640 analyzer were with high precision for the detection of K + ,Na+ and Cl- ,in‐tra‐assay and inter‐assay coefficient of variation of which were both less than 2% .The correlation coefficients of the two analyzers were higher than 0 .975 .Conclusion Cobas8000 analyzer and AU640 analyzer might be with fine comparability for the detection of K+ 、Na+ and Cl- ,both of which could provide accurate detected results .
Texte intégral:
1
Indice:
WPRIM
Type d'étude:
Diagnostic_studies
/
Guideline
langue:
Zh
Texte intégral:
International Journal of Laboratory Medicine
Année:
2015
Type:
Article