Compensation research on energy response for the semiconductor personal dosimeter / 中国医学装备
China Medical Equipment ; (12): 4-6,7, 2015.
Article Dans Chinois | WPRIM | ID: wpr-601936
ABSTRACT

Objective:

To complete experiments of the energy response compensation for three semiconductor personal dosimeters in order to meet the requirements of IEC and National Standard.

Methods:

In Beijing SSDL, the semiconductor detectors were covered by the different thicknesses of Pb and Sn additional filters with several diameter holes to compensate the dosimeters’ energy response.

Results:

At the 50 keV-Cs-137 energy range, the measuring error of dosimeters is met within 30% of energy response.

Conclusion:

By the compensation of combined Pb and Sn filters with a hole, the semiconductor personal dosimeters can be used for radiation monitoring.