The performance evaluation on 3 hemoglobin analyzers / 中国输血杂志
Chinese Journal of Blood Transfusion ; (12): 846-848, 2017.
Article Dans Chinois | WPRIM | ID: wpr-607426
ABSTRACT
Objective To carry out the performance evaluation on 3 hemoglobin analyzers-URIT-12,HemoCue Hb301 and CompoLab TS.Methods Verification of inter-run precision,intra-day precision,and correlation of the 3 analyzers using samples with different concentrations.Results The lower values of inter-run precision of URIT-12,HemoCue Hb301 and CompoLab TSoft were 1.46%,1.45% and 1.32%,respectively,and the median values were 1.43%,0.56% and 1.49%,respectively.The lower values of intra-day precision were 1.86%,1.62% and 1.32% respectively for URIT-12,HemoCue Hb301 and CompoLab TSoft,with the median values of 1.45%,1.95% and 1.78% and the higher values of 1.86%,1.74% and 1.81%.Based on the results of Beckman-Coulter AcT5diff automatic blood cell analyzer,linear regression equation and correlation coefficient was calculated and expressed as y=0.960 2x+3.201 3,R2=0.973 4;y=1.112 1x-6.767 9,R2 =0.990 7;y=1.031 5x+0.634 7,R2=0.980 3,respectively.Conclusion The 3 hemoglobin analyzers have better inter-run precision and intra-day precision as well as good correlation with automatic blood analyzer,and thus can be applied in physical examinations before blood donation.

Texte intégral: Disponible Indice: WPRIM (Pacifique occidental) langue: Chinois Texte intégral: Chinese Journal of Blood Transfusion Année: 2017 Type: Article

Documents relatifs à ce sujet

MEDLINE

...
LILACS

LIS

Texte intégral: Disponible Indice: WPRIM (Pacifique occidental) langue: Chinois Texte intégral: Chinese Journal of Blood Transfusion Année: 2017 Type: Article